Huber

Huber Unichiller P015w-H Recirculating Chiller, -20 to 100 °C

Part no: 3051.0014.01

The Huber Unichiller P015w-H is a water-cooled recirculating chiller with the Pilot ONE controller for -20 to 100 °C. It offers 1.8 - 2.1 kW heating capacity, 1.5 kW cooling capacity at 15 °C and a pump delivering up to 25 l/min at 2.5 bar. Unichillers replace tap-water cooling with a closed circuit at constant temperature, pressure and flow, for removing process heat, cooling technical equipment or supplying a laboratory with cooling water. They are designed for continuous operation at ambient temperatures up to 40 °C.

Temperature range
-20 to 100 °C
Temperature stability
±0.2 K
Controller
Pilot ONE
Safety class
III / FL
See all 38 specifications
Price on request Quoted per configuration and delivery country

We reply to quote requests within one working day.

  • Supplied through our manufacturer partnerships
  • Installation, training and calibration on request
Overview

What this Huber unit gives you, taken from the manufacturer data we hold on file.

  • Available with synthetic refrigerant (R-449A) and natural refrigerant (R-290).
  • Features Environmentally-friendly, economic cooling solution
  • High efficiency
  • Precise temperature control
  • Powerful circulation pump
  • Robust stainless construction
  • Reliable continuously operation with warning functions
  • Easy to fill and empty
  • Adjustable bypass, level indicator with sight glass and digital manometer
  • Pilot ONE controller with 5.7" touchscreen
  • 2 x USB (Host and Device), Ethernet and RS232 interfaces
  • Specifications Synthetic refrigerant (R-449A) Natural refrigerant (R-290) Temperature range -20 to 100 °C -20 to 100 °C Temperature stability ±0.2 K ±0.2 K Controller Pilot ONE Pilot ONE Safety class III / FL III / FL Protection class IP20 IP20 Dimensions (W x D x H) 350 x 496 x 622 mm 350 x 496 x 622 mm Net weight 62 kg 57 kg Heating capacity 1.8 - 2.1 kW 1.8 - 2.1 kW Cooling capacity at 15 °C 1.5 kW 1.5 kW Cooling capacity at 0 °C 1 kW 1 kW Cooling capacity at -10 °C 0.7 kW 0.7 kW Cooling capacity at -20 °C 0.3 kW 0.3 kW Refrigeration machine water-cooled, CFC- and HCFC-free water-cooled, natural refrigerant Refrigerant 1st stage R-449A (A1, H280) R-290 (A3, H220) Refrigerant quantity 1st stage 0.25 kg 0.099 kg Pressure pump, max. 25 l/min; 2.5 bar 25 l/min; 2.5 bar Pump connection G3/4 male G3/4 male Pump speed adjustable no no Pump type B B Pump characteristics Pressure pump, speed not variable, manual bypass integrated Pressure pump, speed not variable, manual bypass integrated External sensor connection Pt100 Pt100 Digital interfaces Ethernet, USB (Host and Device), RS232 Ethernet, USB (Host and Device), RS232 Permissible ambient temperature 5 to 40 °C 5 to 40 °C Sound pressure level 54 dB(A) 54 dB(A) Filling volume min. 3.8 l 3.8 l Expansion volume 1.7 l 1.7 l Cooling water connection G1/2 male G1/2 male Cooling water consumption at 15°C, flow -20°C 36 l/h 36 l/h Cooling water consumption at 15°C, flow -10°C 48 l/h 48 l/h Cooling water consumption at 15°C, flow 0°C 60 l/h 60 l/h Cooling water consumption at 15°C, flow 15°C 68 l/h 68 l/h Cooling water differential pressure min. 1 bar 1 bar Cooling water pressure max. 6 bar 6 bar Power supply 220-240 V 1~ 50 Hz 220-240 V 1~ 50 Hz Current consumption, max. 13.5 A 14.5 A Fuse 16 A 16 A Huber order no. 3051.0014.01 3051.0049.01 European version (220-240 V 1~ 50 Hz or 220-240 V 1~/2~ 50 Hz). Other voltages on request.
  • Price on request. Tell us your application and we send an offer.
Specifications
Temperature range-20 to 100 °C
Temperature stability±0.2 K
ControllerPilot ONE
Safety classIII / FL
Protection classIP20
Dimensions (W x D x H)350 x 496 x 622 mm
Net weight62 kg
Heating capacity1.8 - 2.1 kW
Cooling capacity at 15 °C1.5 kW
Cooling capacity at 0 °C1 kW
Cooling capacity at -10 °C0.7 kW
Cooling capacity at -20 °C0.3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant 1st stageR-449A (A1, H280)
Refrigerant quantity 1st stage0.25 kg
Pressure pump, max.25 l/min; 2.5 bar
Pump connectionG3/4 male
Pump speed adjustableno
Pump typeB
Pump characteristicsPressure pump, speed not variable, manual bypass integrated
External sensor connectionPt100
Digital interfacesEthernet, USB (Host and Device), RS232
Permissible ambient temperature5 to 40 °C
Sound pressure level54 dB(A)
Filling volume min.3.8 l
Expansion volume1.7 l
Cooling water connectionG1/2 male
Cooling water consumption at 15°C, flow -20°C36 l/h
Cooling water consumption at 15°C, flow -10°C48 l/h
Cooling water consumption at 15°C, flow 0°C60 l/h
Cooling water consumption at 15°C, flow 15°C68 l/h
Cooling water differential pressure min.1 bar
Cooling water pressure max.6 bar
Power supply220-240 V 1~ 50 Hz
Current consumption, max.13.5 A
Fuse16 A
Price list noteUnichiller "P" Models: Circulating pumps with a high discharge pressure for applications with high pressure drops.
Huber order no.3051.0014.01
ManufacturerHuber
Part number3051.0014.01

Figures come from the manufacturer. Ask us if you need a value confirmed for your application.

Delivery and support

Lead time depends on the manufacturer and the configuration, so we confirm it with your quote. Delivery, installation, operator training and calibration can all be included, and we will price them separately so you can see what each part costs.

Tell us the country you need it delivered to and we will quote transport and any documentation your institution requires.

Ask about this item

Your quote list

This item is ready to send. Add anything else you need from the related items below and it goes out as one request.

    Not sure which model fits your work?

    Send us your samples, throughput and the standard you work to. We will come back with the model that fits and a written quote.

    Huber Unichiller P015w-H Recirculating Chiller, -20 to 100 °C
    Price on request