What Is GLP in Laboratory? Principles, Rules and GxP

GLP (Good Laboratory Practice) is a quality system for how non-clinical safety studies are planned, performed, monitored, recorded, reported and archived. It exists so that regulators can trust, and reconstruct, the data used to decide whether a medicine, pesticide, chemical or food additive is safe.

GLP does not tell you which test to run or whether your result is scientifically right. It makes sure that every step of the study is documented well enough that an inspector can check it years later.

What is GLP in a laboratory?

In practice, a GLP laboratory (a "test facility" in GLP language) is one that runs studies under a formal set of rules and is inspected by a national GLP monitoring authority. GLP applies to:

  • toxicology studies (acute, repeated dose, reproductive, genotoxicity)
  • ecotoxicology and environmental fate studies
  • physical-chemical property testing of test items
  • residue studies for pesticides and veterinary drugs
  • safety pharmacology and other non-clinical studies submitted to regulators

GLP does not apply to basic research, routine quality control of products (that is GMP), or studies in humans (that is GCP).

Where GLP comes from

GLP rules were introduced in the 1970s after US FDA inspections found serious problems in toxicology data submitted by some contract laboratories, including studies that could not be reconstructed from the records.

  • United States: the FDA issued 21 CFR Part 58 in 1978 (in force from 1979). The EPA has equivalent rules in 40 CFR Part 160 (pesticides) and 40 CFR Part 792 (industrial chemicals).
  • OECD: the OECD Principles of GLP were adopted in 1981 together with the Council Decision on Mutual Acceptance of Data (MAD), and revised in 1997. Under MAD, a GLP study done in one participating country is accepted in the others, so it does not have to be repeated.
  • European Union: Directive 2004/10/EC requires the OECD principles for regulatory safety studies, and Directive 2004/9/EC sets how member states inspect and verify GLP compliance.

The OECD Principles of GLP

The OECD principles are organised in ten sections. They are the core of every national GLP programme.

OECD GLP section What it requires in practice
1. Test facility organisation and personnel Defined management responsibilities, a Study Director for each study, qualified and trained staff with records of training
2. Quality assurance programme An independent QA unit that inspects studies and facilities and audits final reports
3. Facilities Suitable size and separation of activities, so test systems, test items and archives do not mix or contaminate each other
4. Apparatus, material and reagents Equipment that is fit for purpose, inspected, cleaned, maintained and calibrated per SOP; labelled reagents with expiry dates
5. Test systems Proper housing, care and records for physical, chemical and biological test systems (cells, animals, plants)
6. Test and reference items Receipt, identity, characterisation, storage, stability and handling records
7. Standard operating procedures Written, approved SOPs for all routine activities, available where the work is done
8. Performance of the study An approved study plan (protocol) before the study starts; raw data recorded promptly and accurately; changes justified and signed
9. Reporting of study results A final report signed by the Study Director, with a QA statement
10. Storage and retention of records and materials Secure archives with a responsible archivist; records, specimens and samples kept for the required period

GLP laboratory practices: what changes day to day

If you move from a research lab to a GLP lab, these are the habits that change most.

  • Everything starts with a plan. No study work begins before the study plan is approved and signed by the Study Director.
  • Raw data are recorded directly, promptly and legibly, with date and signature. Corrections never hide the original entry and are explained.
  • Every instrument has a logbook: use, cleaning, maintenance, calibration, faults and repairs.
  • Every reagent and solution is labelled with identity, concentration, storage conditions and expiry date.
  • SOPs describe how things are done, and deviations from SOPs or the study plan are documented.
  • Computerised systems are validated, access is controlled, and electronic records keep an audit trail.
  • Specimens, samples and records go to a controlled archive at the end of the study.

Many labs that are not formally GLP-certified still adopt these practices, because they make results easier to defend in audits and customer reviews.

GLP and laboratory equipment

Equipment is where GLP meets the bench. 21 CFR 58.61 requires equipment of appropriate design and adequate capacity, and 58.63 requires that equipment is adequately inspected, cleaned, maintained, tested, calibrated and/or standardised, with written SOPs and records of each operation. The OECD principles say the same in section 4.

For each instrument used in a GLP study, an inspector will usually ask for:

  • proof that it was installed correctly and works to specification (IQ and OQ)
  • evidence that it performs correctly for your method (PQ or equivalent checks)
  • calibration certificates traceable to national or international standards
  • a maintenance and calibration schedule, and records that it was followed
  • for software-controlled instruments, validation of the computerised system and an audit trail

Temperature-controlled equipment (ovens, incubators, climate chambers, freezers) also needs temperature mapping or monitoring records, because storage conditions of test items and specimens are part of the study record.

GLP vs GMP vs GCP: what is GxP?

GxP is the umbrella term for "good practice" quality rules in regulated life sciences. The "x" stands for the area.

Guideline Full name Covers Main question it answers
GLP Good Laboratory Practice Non-clinical safety studies Can this study be trusted and reconstructed?
GMP Good Manufacturing Practice Production and quality control of medicines and other regulated products Is every batch made and tested consistently to its specification?
GCP Good Clinical Practice Clinical trials in humans Are participants protected and are trial data credible?
GDP Good Distribution Practice Storage and transport of medicines Is product quality kept intact through the supply chain?
GVP Good Pharmacovigilance Practice Safety monitoring after approval Are side effects detected and reported?

A common source of confusion: a pharmaceutical QC laboratory testing batches for release works under GMP, not GLP, even though it is a laboratory. GLP is tied to safety studies, not to product release.

Across all GxP areas, data integrity follows the ALCOA+ principles: data must be Attributable, Legible, Contemporaneous, Original and Accurate, plus Complete, Consistent, Enduring and Available. For electronic records, FDA 21 CFR Part 11 and EU GMP Annex 11 set the requirements, and OECD GLP Advisory Document No. 17 covers computerised systems in GLP.

GLP vs ISO/IEC 17025

ISO/IEC 17025 is an accreditation standard for the competence of testing and calibration laboratories. It focuses on validated methods, measurement uncertainty and traceability for routine tests. GLP focuses on the organisation and documentation of individual studies. A contract lab can hold both: 17025 accreditation for routine analyses, and GLP compliance for regulatory safety studies.

GLP laboratory requirements: a short checklist

  • Test facility management with written responsibilities
  • A Study Director named for each study
  • An independent quality assurance unit
  • Approved SOPs for every routine activity
  • Training records and CVs for all staff
  • Qualified, calibrated and maintained equipment with logbooks
  • Controlled handling of test and reference items
  • Validated computerised systems with audit trails
  • A secure, access-controlled archive
  • Inspection by the national GLP monitoring authority

How BIEL can help

BIEL engineering supplies laboratory equipment from Memmert, BUCHI, IKA, BRAND, Hettich, Leica, Metrohm and other manufacturers to labs in North Macedonia and the Balkans, and supports the equipment side of GLP and GMP compliance. We install and calibrate the instruments we supply, prepare IQ/OQ/PQ qualification documentation, and build turnkey laboratories with eLaborator lab furniture. See our services, climate chambers, incubators, ovens and analytical instruments.

FAQ

What does GLP stand for in a laboratory?

Good Laboratory Practice: a set of principles for planning, performing, monitoring, recording, reporting and archiving non-clinical safety studies submitted to regulators.

What is the difference between GLP and GMP laboratory testing regulations?

GLP covers non-clinical safety studies that support product approval. GMP covers manufacturing and quality control testing of products once they are made. A toxicology study is GLP; a batch release test in a pharma QC lab is GMP.

What is the difference between a GMP and a non-GMP laboratory?

A GMP laboratory works under a pharmaceutical quality system that is inspected by the medicines authority: qualified equipment, validated methods, controlled documents and data integrity rules. A non-GMP lab may use good practices but its results cannot be used to release regulated products.

Does every laboratory need to follow GLP?

No. GLP is mandatory only for non-clinical safety studies submitted to regulators. Research, teaching, clinical and routine QC labs follow other rules, though many borrow GLP practices such as SOPs, logbooks and calibration records.

What is GxP?

GxP is the collective name for good practice guidelines in regulated industries: GLP, GMP, GCP, GDP, GVP and others. They share the same core ideas of documented procedures, trained people, qualified equipment and reliable data.

What equipment documentation does a GLP lab need?

Installation and operational qualification records, calibration certificates, maintenance and cleaning logs, SOPs for use and calibration, and for software-controlled instruments a computerised system validation with audit trail.

Request a quote for qualified, calibrated equipment for your GLP or GMP lab.

Back to blog